INTERCOMPARISONS OF GA-67 AND I-123 ASSAYS IN UK HOSPITALS, 1996

Citation
Mj. Woods et al., INTERCOMPARISONS OF GA-67 AND I-123 ASSAYS IN UK HOSPITALS, 1996, Applied radiation and isotopes, 49(9-11), 1998, pp. 1449-1452
Citations number
6
Categorie Soggetti
Nuclear Sciences & Tecnology","Radiology,Nuclear Medicine & Medical Imaging","Chemistry Inorganic & Nuclear
Journal title
Applied radiation and isotopes
ISSN journal
09698043 → ACNP
Volume
49
Issue
9-11
Year of publication
1998
Pages
1449 - 1452
Database
ISI
SICI code
0969-8043(1998)49:9-11<1449:IOGAIA>2.0.ZU;2-U
Abstract
As part of its ongoing support to the National Measurement System in t he UK, the National Physical Laboratory (NPL) conducted two intercompa rison exercises in 1996. The participants, hospital physicists, were r equired to assay samples of Ga-67 and I-123 using radionuclide calibra tors. The analysis of the results presented here, whilst encouraging i n terms of overall level of quality compared with previous exercises, identified several areas of concern. The results are discussed, conclu sions drawn and recommendations made. (C) 1998 The National Physics La boratory. Published by Elsevier Science Ltd. All rights reserved.