ANALYSIS OF SELECTED GASEOUS ORGANIC MICRO-POLLUTANTS IN AIR BY MEANSOF SOLID-PHASE MICROEXTRACTION GAS-CHROMATOGRAPHY MASS-SPECTROMETRY, CALIBRATION PROBLEMS
J. Namiesnik et al., ANALYSIS OF SELECTED GASEOUS ORGANIC MICRO-POLLUTANTS IN AIR BY MEANSOF SOLID-PHASE MICROEXTRACTION GAS-CHROMATOGRAPHY MASS-SPECTROMETRY, CALIBRATION PROBLEMS, Analusis, 26(4), 1998, pp. 170-174
In this work, problems related to calibration of solid phase microextr
action (SPME) device were described. The effects of humidity and tempe
rature on organic pollutants sampling from gaseous matrices on a 100 m
u m FDMS coated SPME fiber were studied. Standard gaseous mixtures use
d in these studies were generated by the permeation method and humidif
ied to known levels of relative humidity. The SPME-GC-MS based procedu
re was used to determine organic vapours (carbon tetrachloride, toluen
e, chlorobenzene, p-xylene and n-decane) in a laboratory air.