GALLIUM-ARSENIDE AC-COUPLED MICROSTRIP DETECTORS WITH INTEGRATED RESISTORS

Citation
Rl. Bates et al., GALLIUM-ARSENIDE AC-COUPLED MICROSTRIP DETECTORS WITH INTEGRATED RESISTORS, Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 410(1), 1998, pp. 19-25
Citations number
13
Categorie Soggetti
Nuclear Sciences & Tecnology","Physics, Particles & Fields","Instument & Instrumentation",Spectroscopy
ISSN journal
01689002
Volume
410
Issue
1
Year of publication
1998
Pages
19 - 25
Database
ISI
SICI code
0168-9002(1998)410:1<19:GAMDWI>2.0.ZU;2-D
Abstract
GaAs microstrip detector prototypes for the ATLAS forward tracker have been fabricated and tested. They feature integrated capacitors for AC -coupling and a novel biasing technique based on the high-resistivity substrate material with ohmic contacts. The strips are wedge shaped an d the detector has a 5.5 cm long off-axis edge cut. The detector has b een tested with the LBIC binary read-out chip in a beam test performed at the CERN SPS with 120 GeV pi(-), showing a 97% detection efficienc y. (C) 1998 Elsevier Science B.V. All rights reserved.