ANEURYSM CLIPS - ASSESSMENT OF MAGNETIC-FIELD INTERACTION ASSOCIATED WITH A 0.2-T EXTREMITY MR SYSTEM

Citation
Fg. Shellock et Jv. Crues, ANEURYSM CLIPS - ASSESSMENT OF MAGNETIC-FIELD INTERACTION ASSOCIATED WITH A 0.2-T EXTREMITY MR SYSTEM, Radiology, 208(2), 1998, pp. 407-409
Citations number
20
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging
Journal title
ISSN journal
00338419
Volume
208
Issue
2
Year of publication
1998
Pages
407 - 409
Database
ISI
SICI code
0033-8419(1998)208:2<407:AC-AOM>2.0.ZU;2-V
Abstract
Twenty-two different aneurysm clips were tested by means of the deflec tion angle test for magnetic field interaction associated with a 0.2-T extremity magnetic resonance (MR) system. At the opening of the magne t bore and 5 inches (13 cm) and 10 inches (25 cm) away from the magnet bore, respectively, deflection angles were 14 degrees, 2 degrees and 0 degrees for one clip and were 17 degrees, 4 degrees, and 0 degrees f or another, and were 0 degrees for the remaining 20 clips. Because the patient's head does not enter the magnet extremity MR imaging of pati ents with the aneurysm clips evaluated in this study should be safe.