Certified reference materials (CRMs) which can be used to analyze anio
ns in aqueous solution were prepared. Using ion chromatography, highly
concentrated CRM solutions containing seven different anions (F-, Cl-
, NO2-, NO3-, Br-, PO43-, and SO42-, 100 mg kg(-1) each) could be dire
ctly separated without making any diluted solutions. To obtain optimiz
ed separation conditions in this highly concentrated sample solution,
a number of experiments including change of eluent composition and flo
w-rate of the eluents and injection volumes were attempted. The best r
esults were observed when a mixture of 2.0 mM Na2CO3 and 2.5 mM NaHCO3
as an eluent was used with conditions of 0.5 ml min(-1) as flow-rate
and a 5 mu l sample loop. The resolution value reached to 1.5, which i
s sufficient for quantitative analysis and all of seven different anio
ns were perfectly separated within 15 min. (C) 1998 Elsevier Science B
.V. All rights reserved.