R. Escobar et al., DETERMINATION OF CR(III) IN URINE, BLOOD-SERUM AND HAIR USING FLOW-INJECTION CHEMILUMINESCENCE ANALYSIS, Fresenius' journal of analytical chemistry, 361(5), 1998, pp. 509-511
A flow injection (FI) chemiluminescence method for the determination o
f Cr(III) in blood serum, urine and hair samples is reported. It is ba
sed on the chromium-catalyzed light emission from the luminol oxidatio
n by hydrogen peroxide. The apparatus consists of an FI system with a
flow cell formed by a coiled transparent tube suitable for chemilumine
scence detection. The specificity of the method is achieved in presenc
e of EDTA. The detection limit under optimum conditions is 0.01 mu g L
-1 of Cr(III). Precision and accuracy were evaluated by determining Cr
(III) concentrations in urine standards from the National Institute of
Standard and Technology (NIST).