MEASUREMENT OF ABSOLUTE CONCENTRATION AND VIABILITY OF CD34(-BLOOD AND CORD-BLOOD PRODUCTS USING FLUORESCENT BEADS AND CYANINE NUCLEIC-ACIDDYES() CELLS IN CORD)
W. Hubl et al., MEASUREMENT OF ABSOLUTE CONCENTRATION AND VIABILITY OF CD34(-BLOOD AND CORD-BLOOD PRODUCTS USING FLUORESCENT BEADS AND CYANINE NUCLEIC-ACIDDYES() CELLS IN CORD), Cytometry, 34(3), 1998, pp. 121-127
Conventional flow cytometric methods for CD34(+) cell counting may he
affected by the high number of nucleated red blood cells or nonviable
cells in cord blood and its products. We developed a simple Row cytome
tric no-wash procedure that avoids these shortcomings because it provi
des absolute CD34(+) cell counts and assesses cell viability, Samples
were incubated with phycoerythrin (PE)-labeled anti-CD34 (Becton Dicki
nson Immunocytometry Systems [BD], San lose, CA) and peridinin chlorop
hyll protein (PerCP)-labeled anti-CD45 (BD) in bead-containing TRUCOUN
T tubes (BD), After red cell lysis with a fixative-free reagent, the i
mpermeant nucleic acid dye YO-PRO-1 (Molecular Probes, Eugene, OR) was
added and samples were analyzed on a single-laser FACSCalibur (BD), A
comparison with the ProCOUNT progenitor cell assay (BD) in 57 samples
revealed excellent correlation of results (r = 0.98, intercept -0.2 c
ells/mu l, slope 1.01), Precision studies conveyed coefficients of var
iation of 6.4 and 8.9% at concentrations of 35 and 16 CD34(+) cells/mu
l, respectively, In untreated and leukocyte-enriched cord blood 4.5 /- 3.8% of CD34(+) cells were stained by YO-PRO-1, representing apopto
tic or necrotic cells. In post-thawing cryopreserved samples this numb
er increased to 10.4 +/- 5.5%, Isotype controls showed very low blank
values of viable cells (0.1 +/- 0.4 cells/mu l, maximum 2.4) and seeme
d unnecessary. We found no washing-related alteration of results in 35
samples, indicating that the method may also be performed with cell w
ashing, Replacing YO-PRO-1 with TO-PRO-3 facilitated four-color analys
is of subpopulations of viable CD34(+) cells on a FACSCalibur equipped
with a second (diode) laser, We found the proposed method to be a rap
id, efficient, and flexible procedure that improved validity of CD34() cell counts. (C) 1998 Wiley-Liss, Inc.