H. Tabata et al., A PARALLEL COMPARISON OF 3 AUTOMATED MULTICHANNEL BLOOD-CELL COUNTINGSYSTEMS FOR ANALYSIS OF BLOOD OF LABORATORY-ANIMALS, Comparative haematology international, 8(2), 1998, pp. 66-71
A parallel evaluation was performed on three automated haematology ana
lysers, the Sysmex K-4500 and Coulter Counter S-PLUS STKR impedance an
alysers, and the Techinicon H1E flow cytometry analyser. The same blo
od samples from animals from three different species were analysed on
the same day. An analytical comparison of haemograms from healthy monk
eys, dogs, rats and phlebotomised rats was made from paired blood samp
les anticoagulated with dipotassium ethylenediaminetetra-acetate. Each
instrument was calibrated with commercially available material based
on human blood products. The precision of each system was good, but wh
ereas the Coulter Counter S-PLUS STKR had better precision for white b
lood cell counts, platelet counts showed greater variability. When com
pared to spun packed cell volume (PCV), or to each of the other instru
ments, the Coulter system consistently gave lower haematocrit values.
The magnitude of relative bias of spun PCV values was about -7% in mon
key, -5% in dogs and -10% in rats. It was deemed necessary with this i
nstrument to make adjustments to the calibration, especially for rats.
The Coulter also gave consistently higher white blood cells (WBC) cou
nts in monkeys, and lower platelet counts in rats compared to the othe
r two instruments. The biases may be due to inherent physical differen
ces between the analytical methods and/or the calibration techniques.
With few exceptions, each instrument provides reliable results for all
major animal species encountered in routine experimental and toxicolo
gical haematology.