Wnp. Lee et al., CALIBRATION OF ISOTOPE RATIO MASS-SPECTROMETRY WORKING STANDARD FOR H-2 H-1 RATIO ANALYSIS/, Journal of mass spectrometry, 33(7), 1998, pp. 627-630
APE is the atom percent excess. It was found that the determined r(WR)
had a precision of +/-1% Most of the variation came from sample handl
ing, which includes the serial dilution of the weighed deuterium oxide
, the reduction of water to hydrogen gas and the introduction of the h
ydrogen gas into the mass spectrometer. The use of this calibration pr
ocedure allows the direct determination of the isotope ratio r(SA), an
d APE from a standard curve over a wide range of enrichment. Unlike th
e reference standards V-SMOW and SLAP, the highly purified deuterium o
xide is non-exhaustible and readily available. The highly purified deu
terium oxide should be used in place of secondary standards to calibra
te working references for H-2/H-1 analysis in biological studies. (C)
1998 John Wiley & Sons, Ltd.