We characterized the surface structure of 21 retrieved femoral ICLH cu
ps, as an indirect measure of surface damage, with an optical scanner
specially developed for surface roughness measurements. 3 unused cups
were measured and served as controls. The mean insertion time was 10 (
5-15) years. 5 areas on each cup were measured, 1 on the top and 4 at
the edge. The mean value of 3 different, commonly used, surface roughn
ess parameters was calculated for the edge measurements. We found no c
orrelation between the surface topography and clinical patient data su
ch as age, body-weight, head-shaft angle, implantation time and wear o
f the polyethylene socket.