E. Beirnaert et al., DESIGN AND EVALUATION OF AN IN-HOUSE HIV-1 (GROUP-M AND GROUP-O), SIVMND AND SIVCPZ ANTIGEN CAPTURE ASSAY, Journal of virological methods, 73(1), 1998, pp. 65-70
Citations number
19
Categorie Soggetti
Virology,"Biochemical Research Methods","Biothechnology & Applied Migrobiology
An enzyme-linked immuno-sorbent assay (ELISA) for the detection of hum
an immunodeficiency virus type 1 (HIV-1) and simian immunodeficiency v
irus (SIVcpz/SIVmnd) antigens was designed using immunoreagents from n
aturally infected individuals, and compared to the commercially availa
ble Vironostika HIV-1 Antigen Microelisa System (Organon Teknika). The
in-house assay proved to be specific for HIV-1 isolates belonging to
group M (A-H) and group O and for SIVcpz and SIVmnd isolates, but was
less sensitive than the Vironostika HIV-1 Antigen Microelisa System, e
xcept for SIVmnd. For the strains belonging to HIV-2, SIVmac and SIVag
m, the in-house assay could not detect antigen to an appreciable degre
e. This study shows that a considerably less expensive but sufficientl
y accurate HIV-1 antigen capture assay can be developed to monitor HIV
-1 (group M and O), SIVcpz and SIVmnd, antigen in the supernatants of
virus cultures. (C) 1998 Elsevier Science B.V. All rights reserved.