Nv. Kashirin et al., SYSTEM OF ACTIVE REGULATION OF THE TEMPERATURE AND TEMPERATURE-GRADIENT IN NUCLEAR-MAGNETIC-RESONANCE SENSORS, Instruments and experimental techniques, 41(3), 1998, pp. 443-446
An active system of temperature regulation and temperature gradient in
a sample on the example NMR sensor is described in this work. By supp
orting the given average temperature, the system of temperature-stabil
ization allows one to set and supervise the size and direction of the
temperature gradient along a sample. The temperature of a sample is ma
intained with on accuracy of about +/-0.1 degrees C in a temperature i
nterval of -150 to +200 degrees C with a longitudinal temperature grad
ient no more than +0.2 degrees C. The value of the temperature gradien
t is established in the range of 0 degrees to 10 degrees C/cm in both
directions.