SUBSTRATE-LESS, SPARK-FREE MICRO-STRIP GAS COUNTERS

Citation
R. Bellazzini et al., SUBSTRATE-LESS, SPARK-FREE MICRO-STRIP GAS COUNTERS, Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 409(1-3), 1998, pp. 14-19
Citations number
14
Categorie Soggetti
Nuclear Sciences & Tecnology","Physics, Particles & Fields","Instument & Instrumentation",Spectroscopy
ISSN journal
01689002
Volume
409
Issue
1-3
Year of publication
1998
Pages
14 - 19
Database
ISI
SICI code
0168-9002(1998)409:1-3<14:SSMGC>2.0.ZU;2-F
Abstract
We review recent work involving micro-strip gas counters with ''advanc ed passivated'' cathode strips. We present results from tests of a new variation of the MSGC, the planar micro-gap counter (PMGC), with very small (similar to 10 mu m) anode-cathode gap. Gains of up to 3 x 10(4 ) were achieved and gain variations due to charging effects were less than 10% using an ordinary (uncoated) bore-silicate glass substrate. T he PMGC showed no reduction in gain when subjected to an X-ray flux of 4 x 10(5) Hz/mm(2) and survived exposure to alpha particles equivalen t to 75 days' running at LHC with no signs of strip damage. (C) 1998 E lsevier Science B.V. All rights reserved.