FIRST EXPERIMENTAL RESULTS OF A NEW GAMMA-RAY DETECTOR BASED ON A SILICON DRIFT CHAMBER COUPLED TO A SCINTILLATOR

Citation
C. Fiorini et al., FIRST EXPERIMENTAL RESULTS OF A NEW GAMMA-RAY DETECTOR BASED ON A SILICON DRIFT CHAMBER COUPLED TO A SCINTILLATOR, Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 409(1-3), 1998, pp. 395-399
Citations number
11
Categorie Soggetti
Nuclear Sciences & Tecnology","Physics, Particles & Fields","Instument & Instrumentation",Spectroscopy
ISSN journal
01689002
Volume
409
Issue
1-3
Year of publication
1998
Pages
395 - 399
Database
ISI
SICI code
0168-9002(1998)409:1-3<395:FEROAN>2.0.ZU;2-R
Abstract
A gamma-ray detector consisting in a CsI(Tl) scintillator coupled to a Silicon Drift Chamber (SDC), has been realized and tested. Experiment al results, obtained with a prototype based on a scintillator of 3 mm diameter and 5 mm thickness and on a SDC of the same diameter with on- chip electronics, are reported here. The peculiar noise characteristic s of the SDC allow this gamma-ray detector to operate on a wide energy range with high resolution. An energy threshold of about 10 keV at ro om temperature and of few keV at 0 degrees C has been measured, At 662 keV the detector shows an energy resolution of 4.34% FWHM at room tem perature, to our knowledge the best result ever obtained with a scinti llator. (C) 1998 Elsevier Science B.V. All rights reserved.