DETERMINING DETECTOR REQUIREMENTS FOR MEDICAL IMAGING APPLICATIONS

Citation
Nh. Clinthorne et al., DETERMINING DETECTOR REQUIREMENTS FOR MEDICAL IMAGING APPLICATIONS, Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 409(1-3), 1998, pp. 501-507
Citations number
9
Categorie Soggetti
Nuclear Sciences & Tecnology","Physics, Particles & Fields","Instument & Instrumentation",Spectroscopy
ISSN journal
01689002
Volume
409
Issue
1-3
Year of publication
1998
Pages
501 - 507
Database
ISI
SICI code
0168-9002(1998)409:1-3<501:DDRFMI>2.0.ZU;2-H
Abstract
The Cramer-Rao lower bound on estimator variance is used as a tool for assessing radiation detector requirements for nuclear medicine imagin g. The classical bound for unbiased estimators is reviewed and its rel ationship to the well-known propagation-of-error formula is demonstrat ed. The uniform Cramer-Rao bound is described for situations where unb iased estimation is impractical such as in tomographic imaging. The bo unds are used to evaluate the necessary detector requirements for a Cs I(Tl)/Si photodiode scintillation camera, and to compare the performan ce of a Compton-scatter camera to that of a parallel-hole collimator/A nger camera system. (C) 1998 Published by Elsevier Science B,V, All ri ghts reserved.