HIGH-SPATIAL-RESOLUTION AND REAL-TIME MEDICAL IMAGING USING A HIGH-SENSITIVITY HARPICON CAMERA

Citation
K. Umetani et al., HIGH-SPATIAL-RESOLUTION AND REAL-TIME MEDICAL IMAGING USING A HIGH-SENSITIVITY HARPICON CAMERA, Journal of synchrotron radiation, 5, 1998, pp. 1130-1132
Citations number
11
Categorie Soggetti
Instument & Instrumentation","Physics, Applied",Optics
ISSN journal
09090495
Volume
5
Year of publication
1998
Part
3
Pages
1130 - 1132
Database
ISI
SICI code
0909-0495(1998)5:<1130:HARMIU>2.0.ZU;2-V
Abstract
A HARPICON(TM) camera has been applied to a digital angiography system with fluorescent-screen optical-lens coupling. It uses avalanche mult iplication in the photoconductive layer for high-sensitivity imaging. The limiting spatial resolutions in the 1050 scanning-line mode of the camera are about 30 and 50 mu m at input held sizes of 20 x 20 and 50 x 50 mm on the screen, respectively. For high-speed imaging, the 525 scanning-line mode at a rate of 60 images s(-1) can be selected. High- quality images of coronary arteries in dogs were obtained by intra-aor tic coronary angiography and superselective coronary angiography using a single-energy X-ray above the iodine K-edge energy.