A laser-heated thermoluminescence dosimetry (LHTLD) system was-tested
for albedo response to mono-energetic neutrons ranging in energy from
200 keV to 2.8 MeV. For neutron response measurements, pairs of magnes
ium borate ((MgB4O7)-B-10/11:Tm) and lithium fluoride ((LiF)-Li-6/7:Mg
,Ti) phosphor chips were used. In general, the LHTLD system based on (
MgB4O7)-B-10/11:Tm shows response similar to the LiF system, that is a
decreasing response ratio (R = dose read/dose applied) with increasin
g neutron energy approaching a response of 0.07 for higher energies. I
n addition, the neutron responses consistently indicates a higher resp
onse for the LiF system than the LHTLD system by about 22%, possibly d
ue in part to the internal shielding of non-albedo thermal neutrons in
the LHTLD badge. Published by Elsevier Science Ltd.