DIGITAL SELENIUM RADIOGRAPHY - COMPARISON OF IMAGE QUALITY OF THORACIC X-RAYS IN NORMAL AND REDUCED IMAGE FORMATS, BASED ON STRUCTURAL ANATOMIC DETAILS
M. Zahringer et al., DIGITAL SELENIUM RADIOGRAPHY - COMPARISON OF IMAGE QUALITY OF THORACIC X-RAYS IN NORMAL AND REDUCED IMAGE FORMATS, BASED ON STRUCTURAL ANATOMIC DETAILS, RoFo. Fortschritte auf dem Gebiete der Rontgenstrahlen und der neuenbildgebenden Verfahren, 169(1), 1998, pp. 33-37
Citations number
16
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging
Journal title
RoFo. Fortschritte auf dem Gebiete der Rontgenstrahlen und der neuenbildgebenden Verfahren
Purpose: Is there a loss of information when using selenium radiograph
ic images displayed with reduced image format compared to full format?
Method: Digital selenium radiographic chest images in two planes were
obtained in 35 patients for medical reaons. The digital data sets of
each patient were separately displayed in full format on two different
films and were printed in reduced format on one film. The format was
reduced to 61% of the full format using an acquisition matrix of 2166
x 2488 pixels, an image display matrix of 4000 x 5000 pixels and a fil
m format of 43 x 49 cm. All images were anonymously evaluated by four
independent readers using a questionnaire concerning anatomic structur
es. Results: Format reduction did not result in a loss of information
in diagnostically relevant anatomic details. Conclusions: Displaying d
igital selenium radiographic data sets of thoracic view in two planes
on one film presumably does not lead to a loss of diagnostic informati
on. This procedure may help to reduce film costs.