L. Thomachot et al., PRESERVATION OF HUMIDITY AND HEAT OF RESPIRATORY GASES IN SPONTANEOUSLY BREATHING, TRACHEOSTOMIZED PATIENTS, Acta anaesthesiologica Scandinavica, 42(7), 1998, pp. 841-844
Background: Ventilation with endotracheal intubation bypasses the uppe
r airway and the normal heat and moisture exchanging process of inspir
ed gases. A continuous loss of moisture and heat occurs and predispose
s patients to serious airway damage. We therefore prospectively studie
d one heated humidifier system, one cold humidifier system and one hea
t and moisture exchanger in spontaneously breathing, tracheostomized i
ntensive care unit patients to determine the ability to preserve patie
nts' heat and water. Methods: Following a randomized order, 10 patient
s were spontaneously ventilated for 24-h periods with the Nam 35(R) hu
midifier (Europe Medical, France, a cold water humidifier), the heat a
nd moisture exchanger Trach-Vent(R) (Gibeck, Sweden), and the Aerodyne
(R) humidifer (Kendall, USA). in each patient, during the inspiration
phase, the following measurements were performed: mean values of tempe
rature and relative humidity of inspired gases. The absolute humidity
was calculated. Values were obtained in each patients after 40 min and
24 h. Results: The Trach-Vent(R) filter and the Aerodyne(R) humidifie
r had better humidification and thermic capacities than the Nam 35(R)
humidifier (P<0.001). With the Nam 35(R) humidifier, no patient had te
mperature of inspired gas >29 degrees C Concerning absolute humidity o
f inspired gases, the Nam 35(R) humidifier achieved a lower performanc
e than the other two tested systems (P<0.001). Conclusion: In spontane
ously breathing, tracheostomized intensive care unit patients, the Tra
ch-Vent(R) heat and moisture exchanger and the Aerodyne(R) heated syst
em achieved satisfactorily preservation of heat and humidity of inspir
ed gases.