Rp. Rodgers et al., 5.6 TESLA FOURIER-TRANSFORM ION-CYCLOTRON RESONANCE MASS-SPECTROMETERFOR ANALYSIS OF VOLATILE COMPLEX MIXTUUES, Review of scientific instruments, 69(6), 1998, pp. 2278-2284
An all-glass healed inlet coupled to a homebuilt 5.6 T Fourier transfo
rm ion cyclotron resonance mass spectrometer generates stable mass spe
ctra from a volatile mixture for 2 h from as little as 1 mu L of injec
ted sample. A dynamic range of similar to 500 and ultrahigh mass resol
ving power, m/Delta m(50%)approximate to 320 000, are shown for a seri
es of test compounds. A raw diesel fuel sample (with hundreds of compo
nents over a wide boiling point range) mass spectrum gave similar to 5
00 resolved peaks over a range, 90 less than or equal to m/z<300, from
which molecular formulas could be assigned from accurate (less than o
r equal to+/-0.8 ppm) mass measurements. (C) 1998 American Institute o
f Physics. [S0034-6748(98)04406-2].