MAGNETIC MEASUREMENTS ON THE TCV TOKAMAK

Citation
Jm. Moret et al., MAGNETIC MEASUREMENTS ON THE TCV TOKAMAK, Review of scientific instruments, 69(6), 1998, pp. 2333-2348
Citations number
6
Categorie Soggetti
Physics, Applied","Instument & Instrumentation
ISSN journal
00346748
Volume
69
Issue
6
Year of publication
1998
Pages
2333 - 2348
Database
ISI
SICI code
0034-6748(1998)69:6<2333:MMOTTT>2.0.ZU;2-5
Abstract
The TCV Tokamak was designed to create a large variety of plasma shape s. Such a large flexibility requires high precision magnetic measureme nts with a good spatial coverage. This article gives a detailed descri ption of the magnetic sensor geometry, fabrication, calibration, the a ssociated electronics, and the diagnostic operation and monitoring. A substantial effort has been made to quantify the precision in the meas urements and a novel method has been developed to derive corrections i n the sensor position and calibration which optimise the consistency o f the entire measurement set. Accuracy of 0.5 mWb in the poloidal flux and 1 mT in the magnetic field with a position error of a few mm have been achieved. (C) 1998 American Institute of Physics. [S0034-6748(98 )02006-1].