CLOSED-LOOP CONTROLLED ADMINISTRATION OF PROPOFOL USING BISPECTRAL ANALYSIS

Citation
E. Mortier et al., CLOSED-LOOP CONTROLLED ADMINISTRATION OF PROPOFOL USING BISPECTRAL ANALYSIS, Anaesthesia, 53(8), 1998, pp. 749-754
Citations number
32
Categorie Soggetti
Anesthesiology
Journal title
ISSN journal
00032409
Volume
53
Issue
8
Year of publication
1998
Pages
749 - 754
Database
ISI
SICI code
0003-2409(1998)53:8<749:CCAOPU>2.0.ZU;2-A
Abstract
Ten patients, undergoing elective orthopaedic surgery under spinal ana esthesia, were sedated with propofol using a closed-loop feedback cont rol system. The bispectral index (BIS), a new processed EEG parameter, was used as control variable. Propofol administration was controlled by a patient individualised adaptive model-based controller incorporat ing target-controlled infusion technology combined with a pharmacokine tic-dynamic model. This feedback control system for propofol administr ation proved to be adequate and safe. BIS was found to be well suited as control variable.