THERMOLUMINESCENT RESPONSE AND RELATIVE EFFICIENCY OF TLD-100 EXPOSEDTO LOW-ENERGY X-RAYS

Citation
I. Gamboadebuen et al., THERMOLUMINESCENT RESPONSE AND RELATIVE EFFICIENCY OF TLD-100 EXPOSEDTO LOW-ENERGY X-RAYS, Physics in medicine and biology, 43(8), 1998, pp. 2073-2083
Citations number
27
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging
ISSN journal
00319155
Volume
43
Issue
8
Year of publication
1998
Pages
2073 - 2083
Database
ISI
SICI code
0031-9155(1998)43:8<2073:TRAREO>2.0.ZU;2-R
Abstract
The dose-response of LiF:Mg,Ti (TLD-100) exposed to 15 and 35 kVp (8.0 +/- 0.1 and 8.1 +/- 0.1 keV effective energy respectively) x-rays and Co-60 gamma-rays has been measured in the dose interval from(1.2-5.4) x 10(3) Gy for x-rays, and from 0.14 to 850 Gy for gamma-rays. In both cases the total TL signal and glow curve peaks 3 to 9 show supralinea rity. The supralinearity function f(D) is similar for both x-ray beams , except for peak 8, where a 30% difference is observed. The maxima of f(D) for the total TL signal and peaks 5 to 8 are 2.1, 1.7, 6.4, 3.3 and 7.5 respectively for 8.1 keV x-rays and 3.7, 3.1, 13.6, 9.9 and 11 .0 for gamma-rays. The measured relative efficiencies for x-rays with respect to Co-60, f,, th, total TL signal and peaks 5 and 7, were 1.04 , 0.97 and 3.2 respectively.