I. Gamboadebuen et al., THERMOLUMINESCENT RESPONSE AND RELATIVE EFFICIENCY OF TLD-100 EXPOSEDTO LOW-ENERGY X-RAYS, Physics in medicine and biology, 43(8), 1998, pp. 2073-2083
The dose-response of LiF:Mg,Ti (TLD-100) exposed to 15 and 35 kVp (8.0
+/- 0.1 and 8.1 +/- 0.1 keV effective energy respectively) x-rays and
Co-60 gamma-rays has been measured in the dose interval from(1.2-5.4)
x 10(3) Gy for x-rays, and from 0.14 to 850 Gy for gamma-rays. In both
cases the total TL signal and glow curve peaks 3 to 9 show supralinea
rity. The supralinearity function f(D) is similar for both x-ray beams
, except for peak 8, where a 30% difference is observed. The maxima of
f(D) for the total TL signal and peaks 5 to 8 are 2.1, 1.7, 6.4, 3.3
and 7.5 respectively for 8.1 keV x-rays and 3.7, 3.1, 13.6, 9.9 and 11
.0 for gamma-rays. The measured relative efficiencies for x-rays with
respect to Co-60, f,, th, total TL signal and peaks 5 and 7, were 1.04
, 0.97 and 3.2 respectively.