N. Petoussihenss et al., CALCULATION OF BACKSCATTER FACTORS FOR DIAGNOSTIC-RADIOLOGY USING MONTE-CARLO METHODS, Physics in medicine and biology, 43(8), 1998, pp. 2237-2250
Backscatter factors were determined for x-ray beams relevant to diagno
stic radiology using Monte Carlo methods. The phantom size considered
most suitable for calibration of dosimeters is a cuboid of 30 x 30 cm(
2) front surface and 15 cm depth. This phantom size also provides a go
od approximation to adult patients. Three different media were studied
: water, PMMA and ICRU tissue; the source geometry was a point source
with varying field size and source-to-phantom distance. The variations
of the backscatter factor with phantom medium and field geometry were
examined. From the obtained data a set of backscatter factors was sel
ected and proposed for adoption as a standard set for the calibration
of dosimeters re? be used to measure diagnostic reference doses.