MICROCHANNEL PLATE ELECTRON MULTIPLIER CALIBRATION USING A DISCRETE DETECTOR ARRAY

Citation
Dj. Narayan et al., MICROCHANNEL PLATE ELECTRON MULTIPLIER CALIBRATION USING A DISCRETE DETECTOR ARRAY, International journal of mass spectrometry and ion processes, 176(3), 1998, pp. 161-166
Citations number
16
Categorie Soggetti
Spectroscopy,"Physics, Atomic, Molecular & Chemical
ISSN journal
13873806
Volume
176
Issue
3
Year of publication
1998
Pages
161 - 166
Database
ISI
SICI code
1387-3806(1998)176:3<161:MPEMCU>2.0.ZU;2-F
Abstract
The performance of a microchannel plate electron multiplier (MCP) is k nown to vary with time and exposure to ions, the atmosphere, and other factors. Clearly the recognition of deterioration in experimental wor k is of great importance. In the present work the performance of an MC P is straightforwardly measured using a detector array integrated on a silicon chip and results presented. The objective of this work is not to accurately quantify the MCP gain distribution but rather to find t he distribution shape to enable straightforward monitoring of variatio ns in gain distribution and identification of conditions under which e ssentially all MCP output pulses are recorded. Thus, isotope ratios of high quality can be measured. This work further demonstrates the vers atility of a discrete detector array developed at Aberystwyth. It enab les straightforward remote monitoring of the MCP performance and is es pecially valuable in space-borne experiments. (C) Elsevier Science B.V .