Jt. Heverhagen et al., PITFALLS WHEN USING A CONTRAST-MEDIUM INJECTOR IN MRI, RoFo. Fortschritte auf dem Gebiete der Rontgenstrahlen und der neuenbildgebenden Verfahren, 169(2), 1998, pp. 198-200
Citations number
10
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging
Journal title
RoFo. Fortschritte auf dem Gebiete der Rontgenstrahlen und der neuenbildgebenden Verfahren
Purpose: Using a power injector to applicate Gd-DTPA we found a contra
st enhancement of the pyelo-caliceal system even in the native studies
, and hence we analysed pitfalls when using power injectors in MRI. Ma
terial and methods: We used a power injector Spectris(R) (MedRad, Maas
tricht, Netherlands). In vitro artifacts were achieved by the mixture
of contrast media and saline solution. We substituted contrast media b
y red water, NaCl by clear water. Results: Using power injectors in MR
I, some pitfalls must be avoided, which can render investigations usel
ess, especially dynamic contrast-enhanced investigations. Conclusion:
In our study we showed an easy way to overcome some pitfalls and use a
power injector in MRI in a diagnostically helpful way. The simple use
of valves inhibits the mixture of contrast media and saline solution.