R. Carpignano et al., CHEMOMETRIC APPROACH TO EVALUATE THE PARAMETERS AFFECTING THE DETERMINATION OF REACTION-RATE CONSTANTS BY ION-TRAP A LASS SPECTROMETRY, Journal of the American Society for Mass Spectrometry, 9(9), 1998, pp. 938-944
The effects of experimental parameters on kinetic measurements perform
ed with an ion trap mass spectrometer have been investigated by a chem
ometric approach using Experimental Design techniques. The selected ex
periment was the determination of the rate constant of the reaction CH
3CO++CH3COCH3 --> CH3CO(H)CH3++CH2CO. In a first step, eight variables
were studied: temperature, acetone pressure, buffer gas pressure, met
hod of ion selection, ionization start mass, ionization time, mass ran
ge, and reaction time. The results show that isolation method and reac
tion time were the only parameters affecting rate constants determinat
ion, although temperature may be considered a ''borderline'' factor. T
he remaining parameters and their interactions do not significantly af
fect the kinetic measurements. A second series of runs was performed t
o check again the effect of reaction time, temperature, and ionization
time in experiments with or without ion selection in order to compare
the results of the present work with those obtained in a similar stud
y on a Fourier transform-ion cyclotron resonance instrument. (C) 1998
American Society for Mass Spectrometry.