Vg. Potapova et al., ELECTROTHERMAL ATOMIC-ABSORPTION DETERMINATION OF SILICON IN SINGLE-CRYSTALS BASED ON ALKALI-METAL HALIDES, Journal of analytical chemistry, 53(8), 1998, pp. 768-771
Conditions for determining silicon in cesium iodide and potassium chlo
ride by electrothermal atomic absorption spectrometry were examined. M
easurements were performed using a Saturn spectrophotometer with a Gra
fit-2 atomizer. It was shown that the sensitivity of silicon determina
tion is maximum when graphite tubes are pretreated with a solution of
zirconyl chloride and ferric nitrate is added to the test solution as
a matrix modifier. The tolerable concentrations of cesium iodide and p
otassium chloride in the test solution are 50 and 20 g/L, respectively
. Procedures were developed for determining 3 x 10(-5) mass % of silic
on in cesium iodide and 2 x 10(-4) mass % of silicon in potassium chlo
ride.