Ventricular Defibrillation and Cardiac Function. Introduction: The eff
ect of implantable defibrillator shocks on cardiac hemodynamics is poo
rly understood. The purpose of this study was to test the hypothesis t
hat ventricular defibrillator shocks adversely effect cardiac hemodyna
mics. Methods and Results: The cardiac index was determined by calcula
ting the mitral valve inflow with transesophogeal Doppler during nonth
oracotomy defibrillator implantation in 17 patients. The cardiac index
was determined before, and immediately, 1 minute, 2 minutes, and 4 mi
nutes after shocks were delivered during defibrillation energy require
ment testing with 27- to 34-, 15-, 10-, 5-, 3-, or 1-J shocks. The car
diac index was also measured at the same time points after 27- to 34-,
and 1-J shocks delivered during the baseline rhythm. The cardiac inde
x decreased from 2.30 +/- 0.40 L/min per m(2) before a 27- to 34-J sho
ck during defibrillation energy requirement testing to 2.14 +/- 0.45 L
/min per m(2) immediately afterwards (P = 0.001). This effect persiste
d for > 4 minutes. An adverse hemodynamic effect of similar magnitude
occurred after 15 J (P = 0.003) and 10-J shocks (P = 0.01), but dissip
ated after 4 minutes and within 2 minutes, respectively. There was a s
ignificant correlation between shock strength and the percent change i
n cardiac index (r = 0.3, P = 0.03). The cardiac index decreased 14% a
fter a 27- to 34-J shock during the baseline rhythm (P < 0.0001). This
effect persisted for < 4 minutes. A 1-J shock during the baseline rhy
thm did not effect the cardiac index. Conclusion: Defibrillator shocks
> 9 J delivered during the baseline rhythm or during defibrillation e
nergy, requirement testing result in a 10% to 15% reduction in cardiac
index, whereas smaller energy shocks do not affect cardiac hemodynami
cs. The duration and extent of the adverse effect are proportional to
the shock strength. Shock strength, and not ventricular fibrillation,
appears to be most responsible for this effect. Therefore, the detrime
ntal hemodynamic effects of high-energy shocks may be avoided when low
-energy defibrillation is used.