RESPONSE FUNCTIONS OF THE 4-PI DETECTOR ORION TO 10-70 MEV NEUTRONS

Citation
Y. Perier et al., RESPONSE FUNCTIONS OF THE 4-PI DETECTOR ORION TO 10-70 MEV NEUTRONS, Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 413(2-3), 1998, pp. 312-320
Citations number
17
Categorie Soggetti
Nuclear Sciences & Tecnology","Physics, Particles & Fields","Instument & Instrumentation",Spectroscopy
ISSN journal
01689002
Volume
413
Issue
2-3
Year of publication
1998
Pages
312 - 320
Database
ISI
SICI code
0168-9002(1998)413:2-3<312:RFOT4D>2.0.ZU;2-S
Abstract
The neutron detection performances of the 4 pi ORION counter were inve stigated with monoenergetic neutron beams delivered by the Louvain-la- Neuve facility. Two responses are supplied by such a counter: a prompt response, arising from the slowing down of neutrons within the liquid scintillator contained in the detector, and a delayed response corres ponding to the subsequent capture of the neutrons by Gd nuclei: with w hich the detector is loaded. Two experimental methods were used, allow ing one to probe the responses for different neutron energies. Despite the large size of the detector, the prompt response is observed to be very similar to that exhibited by small cells. For both responses, a reasonable overall agreement is found between the data and the predict ions of Monte-Carlo simulations. The timing resolution. of the order o f 3 ns, makes this detector suitable for time-of-flight methods in app lications requiring a high detection efficiency while accepting a mode rate energy resolution. (C) 1998 Elsevier Science B.V. All rights rese rved.