PERFORMANCE-CHARACTERISTICS OF A CHEMICAL IMAGING TIME-OF-FLIGHT MASS-SPECTROMETER

Citation
Rm. Braun et al., PERFORMANCE-CHARACTERISTICS OF A CHEMICAL IMAGING TIME-OF-FLIGHT MASS-SPECTROMETER, Rapid communications in mass spectrometry, 12(18), 1998, pp. 1246
Citations number
19
Categorie Soggetti
Spectroscopy,"Chemistry Analytical
ISSN journal
09514198
Volume
12
Issue
18
Year of publication
1998
Database
ISI
SICI code
0951-4198(1998)12:18<1246:POACIT>2.0.ZU;2-P
Abstract
A chemical imaging time-of-flight secondary ion mass spectrometer is d escribed. It consists of a liquid metal ion gun, medium energy resolut ion reflectron mass analyzer, liquid nitrogen cooled sample stage, pre paration chamber and dual stage entry port. Unique features include co mpatibility with laser postionization experiments, large field of view , cryogenic sample handling capability and high incident ion beam curr ent. Instrument performance is illustrated by the characterization of scanning electron microscopy grids, silver and functionalized polystyr ene beads and the postionization of an organic overlayer on a gold sub strate, (C) 1998 John Wiley & Sons, Ltd.