Eo. Samwel et al., AN ALTERNATIVE APPROACH TO VECTOR VIBRATING SAMPLE MAGNETOMETER DETECTION COIL SETUP, Review of scientific instruments, 69(9), 1998, pp. 3204-3209
Vector vibrating sample magnetometers (VSMs) can present problems with
respect to angular dependent calibration and positional dependency wh
en they are used for measurements on thin film samples, which have dim
ensions comparable to or larger than the sample-coil distances. The pr
oblems are due to the fact that in conventional VSMs the sample is rot
ating with respect to the coils, when performing angular dependent mea
surements. In this article a solution is presented based on a setup of
VSM detection coils, whose position is linked to that of the sample.
Together with a newly designed sample holder, the above mentioned prob
lems are prevented or reduced. The vector detection coil system shows
a relatively small error in the determination of the magnetization vec
tor (+/-1% in the absolute value and +/-0.6 degrees in the angle). Fur
thermore, it has a relatively small positional dependency (1% per mm)
combined with a sufficient sensitivity (1 nA m(2) or 1 mu emu at 10 s
time constant) and a capability of using samples up to 10X10 mm(2). Th
e improved sample holder for thin film measurements reduces positional
problems while, at the same time, reducing the background signals of
the holder (to 10 pA m(2) per kA/m or 7.958X10(-10) emu/Oe). (C) 1998
American Institute of Physics. [S0034-6748(98)04008-8]