J. Pawlyta et al., COMMISSIONING OF A QUANTULUS 1220(TM) LIQUID SCINTILLATION BETA-SPECTROMETER FOR MEASURING C-14 AND H-3 AT NATURAL-ABUNDANCE LEVELS, Radiocarbon, 40(1), 1998, pp. 201-209
In 1994, the Gliwice Radiocarbon Laboratory began operating a liquid s
cintillation spectrometry system, consisting of a Quantulus 1220(TM) s
pectrometer and two vacuum rigs for benzene production. This paper des
cribes the procedures used for the benzene synthesis from samples cont
aining <1 g of carbon and in the range 1 to 10 g of carbon. We also pr
esent the Quantulus calibration procedures used in the Gliwice Radioca
rbon Laboratory and NERC Radiocarbon Laboratory, and compare the calib
ration parameters.