Quantitative, vascular T-2 measurements are of interest for applicatio
ns such as MR oximetry, In the situation of a vessel with long T-2 rel
axation times embedded in tissue with relatively short T-2 values, con
tamination of the blood signal from the surrounding tissue can bias T-
2 measurements. Limited data sampling and vessels running obliquely th
rough the imaging slice can cause significant signal contamination, Us
ing a model of these effects to predict the behavior of T-2 measuremen
ts under a range of conditions, a set of parameters that provides the
best combination of measurement accuracy with a signal-to-noise ratio
as high as possible is proposed.