VOLUMETRIC MEASUREMENTS OF BONE-MINERAL DENSITY OF THE LUMBAR SPINE -COMPARISON OF 3 GEOMETRICAL APPROXIMATIONS USING DUAL-ENERGY X-RAY ABSORPTIOMETRY (DXA)
Mf. Schreuder et al., VOLUMETRIC MEASUREMENTS OF BONE-MINERAL DENSITY OF THE LUMBAR SPINE -COMPARISON OF 3 GEOMETRICAL APPROXIMATIONS USING DUAL-ENERGY X-RAY ABSORPTIOMETRY (DXA), Nuclear medicine communications, 19(8), 1998, pp. 727-733
Measurement of bone mineral density using dual-energy X-ray absorptiom
etry (DXA) gives area values (g cm(-2)) rather than true volumetric va
lues (g cm(-3)). To calculate the vertebral volume using planar poster
o-anterior and lateral DXA values, several different geometrical appro
ximations were used: cubic, cylindrical with a circular cross-section
and cylindrical with an elliptical cross-section. The aim of this stud
y was to compare these geometrical approximations with each other and
with a reference standard, defined as the volume found on a computed t
omographic (CT) scan. L2 and L3 were evaluated in a phantom study. Vol
ume approximations by the cube or cylinder with circular cross-section
geometry showed more than a 50% overestimation (range 54-74%). Howeve
r, the elliptical cylinder approach showed very good agreement: 2.1% a
nd 1.2% for L2 and L3, respectively, when compared to the CT volumes.
In addition, we performed four patient studies with both CT and DXA to
evaluate the elliptical cylinder estimate in a clinical setting. For
L2 and L3, the mean relative difference was less than 2%. We conclude
that the elliptical cylinder approach results in the most accurate bon
e volume estimates in both the phantom and patients. ((C) 1998 Lippinc
ott Williams & Wilkins).