A POLY(O-AMINOPHENOL) MODIFIED ELECTRODE AS AN AMPEROMETRIC HYDROGEN-PEROXIDE BIOSENSOR

Citation
Mav. Garcia et al., A POLY(O-AMINOPHENOL) MODIFIED ELECTRODE AS AN AMPEROMETRIC HYDROGEN-PEROXIDE BIOSENSOR, Electrochimica acta, 43(23), 1998, pp. 3533-3539
Citations number
36
Categorie Soggetti
Electrochemistry
Journal title
ISSN journal
00134686
Volume
43
Issue
23
Year of publication
1998
Pages
3533 - 3539
Database
ISI
SICI code
0013-4686(1998)43:23<3533:APMEAA>2.0.ZU;2-#
Abstract
An enzymatic biosensor for hydrogen peroxide based on a horseradish pe roxidase (HRP)-ferrocene carbon paste modified electrode and coated wi th a layer of electrochemically generated poly(o-aminophenol) is repor ted. A linear calibration curve is obtained over the range 1 x 10(-8) M to 1 x 10(-5) M. The biosensor responds to hydrogen peroxide ina few seconds and has a detection limit of 8.5 x 10(-9) M. The response of the biosensor is diffusion controlled at low substrate concentrations. Flow injection assays of hydrogen peroxide at a sampling rate of 150 injections per hour with a relative standard deviation of 0.8% (50 sam ples) are possible. Applicability of the sensor for measurement of hyd rogen peroxide in real samples (milk) was demonstrated. (C) 1998 Elsev ier Science Ltd. All rights reserved.