A MICROANGIOGRAPHIC TECHNIQUE USING SYNCHROTRON-RADIATION TO VISUALIZE DERMAL CIRCULATION IN-VIVO
Citation
K. Ito et al., A MICROANGIOGRAPHIC TECHNIQUE USING SYNCHROTRON-RADIATION TO VISUALIZE DERMAL CIRCULATION IN-VIVO, Plastic and reconstructive surgery, 102(4), 1998, pp. 1128-1133
Categorie Soggetti
Surgery
SICI code
0032-1052(1998)102:4<1128:AMTUST>2.0.ZU;2-T
Abstract
Conventional angiography cannot resolve dermal small vessels with a di
ameter of 200 mu m or less. In vitro microangiography is currently cha
racterized by better spatial resolution than conventional angiography
but does not allow visualization of the blood stream in dermal vessels
in vivo. In this study, we introduce a novel synchrotron radiation mi
croangiographic system for visualizing the structure of and blood flow
in dermal microvessels in vivo repeatedly. We used monochromatic sync
hrotron radiation with an energy just above the k-edge of iodine (33.3
keV) as an x-ray and a high-definition television camera system with
a high-sensitivity image pick-up tube for detection. The 33.3-keV mono
chromatic synchrotron radiation allows detection of a small amount of
iodine, and the high-definition television camera system can resolve s
mall vessels with high-spatial resolution and no loss of sensitivity.
We performed synchrotron radiation angiography of superficial inferior
epigastric arteries and their branches in nine rats, and of the cauda
l artery in 14 rats, and compared angiographic images taken by the cur
rent system with those taken by a conventional angiographic system in
seven rats. With this new microangiographic technique, we could visual
ize small dermal vessels with a diameter as low as 50 mu m. In additio
n, repeated angiograms at baseline and under increasing body temperatu
re could be obtained. This new microangiographic approach is expected
to be very useful for the assessment of dermal circulation in patients
.