Pa. Scott et al., ACCURACY OF REAGENT STRIPS IN DETECTING HYPOGLYCEMIA IN THE EMERGENCYDEPARTMENT, Annals of emergency medicine, 32(3), 1998, pp. 305-309
Study objective: Although reagent strips are commonly used, their reli
ability to estimate blood glucose concentration and guide administrati
on of dextrose solutions in the emergency department environment has n
ot been proved. We determined the accuracy of visually interpreted rea
gent strips (Chemstrip bG, Boehringer Mannheim Corp, Indianapolis, IN)
and their ability to identify hypoglycemic patients in the ED. Method
s: We conducted a prospective, nonrandomized blinded clinical study of
the visual estimation of blood glucose values by ED personnel using C
hemstrip bG reagent strips during a 4-month period. Simultaneously obt
ained blood samples sent for laboratory glucose determination served a
s controls. The study was con ducted at a large university hospital ED
with an urban patient population. A convenience sample of 215 adult E
D patients underwent serum glucose determination with data form comple
tion. No study intervention was tested, although timing of administrat
ion of dextrose solutions, if given, was recorded. Results: Hypoglycem
ia was defined as a glucose concentration less than 60 mg/dL on standa
rd laboratory analysis. Reagent strips identified 28 of 29 of these pa
tients (sensitivity=97%), and 171 of 182 patients without hypoglycemia
(specificity=94%, negative predictive value=99%) compared with contro
l samples. The 1 false-negative reagent strip reading of 80 mg/dL was
obtained from blood stored in a serum separator tube and had a laborat
ory glucose value of 39 mg/dL. Eighty-seven percent of the reagent str
ips were within +/- 60 mg/dL of the control value for the laboratory g
lucose reference range less than 350 mg/dL. Conclusion: Visually inter
preted Chemstrip bG reagent strips provide an acceptable estimation of
blood glucose concentration in the ED and are highly sensitive in det
ecting hypoglycemia.