L. Poller et al., A COMPARISON OF LINEAR AND ORTHOGONAL REGRESSION-ANALYSIS FOR LOCAL INR DETERMINATION IN ECAA COAGULOMETER STUDIES, British Journal of Haematology, 102(4), 1998, pp. 910-917
International sensitivity index calibrations based on the W.H.O, recom
mended method depend on orthogonal regression analysis, As this is not
readily available in statistical packages, comparison has been made w
ith simple linear regression analysis in a study of coagulometer effec
ts on the International Normalized Ratio (INR) at 155 European centres
, Sets of seven lyophilized normal and 20 lyophilized artificially dep
leted abnormal plasmas were provided with five coumarin test plasmas a
nd two European Concerted Action on Anticoagulation reference thrombop
lastins (low International Sensitivity Index (ISI) human and high ISI
rabbit). Local ISI based on the artificially depleted lyophilized plas
mas using conventional orthogonal regression gave good correction for
local coagulometer effects on the human reagent and minimal correction
with the rabbit reagent INR. Results were considerably worse after at
tempts at correction using calibration based on linear regression anal
ysis with both reagents. The results indicate that calibration of coag
ulometer prothrombin time systems using simple linear regression is no
t appropriate.