A METHOD FOR THE INTRINSIC CALIBRATION OF CSI(TL) DETECTORS

Citation
Dv. Kamanin et al., A METHOD FOR THE INTRINSIC CALIBRATION OF CSI(TL) DETECTORS, Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 413(1), 1998, pp. 127-137
Citations number
28
Categorie Soggetti
Nuclear Sciences & Tecnology","Physics, Particles & Fields","Instument & Instrumentation",Spectroscopy
ISSN journal
01689002
Volume
413
Issue
1
Year of publication
1998
Pages
127 - 137
Database
ISI
SICI code
0168-9002(1998)413:1<127:AMFTIC>2.0.ZU;2-H
Abstract
A method for the intrinsic energy calibration of photomultiplier-coupl ed CsI(Tl) detectors is described. A simple empirical model of the sci ntillation light pulse-shape of CsI(Tl) crystals for light charged par ticles has been applied to simulate the particle identification matrix as it follows from the pulse-shape analysis method. The calibration p rocedure for the large-area CsI(TI) detectors of the scintillator shel l of the 4 pi-array FOBOS for ions with Z < 4 at energies below 100 AM eV is based on the energies of the particle punch-through points. (C) 1998 Elsevier Science B.V. All rights reserved.