Z. Bosanac et al., ROTATIONAL DIGITAL SUBTRACTION CAROTID ANGIOGRAPHY - TECHNIQUE AND COMPARISON WITH STATIC DIGITAL SUBTRACTION ANGIOGRAPHY, Clinical Radiology, 53(9), 1998, pp. 682-687
Recent advances in angiography equipment have allowed the development
of rotational digital subtraction carotid angiography, We give the fir
st description of this technique applied to the extracranial carotid c
irculation, using the Phillips Integris V3000 release 11, Forty-three
patients undergoing conventional digital subtraction angiography were
compared with 63 patients who had rotational digital subtraction angio
graphy for examination of the extracranial carotid circulation. The pa
rameters compared were: total procedure time, volume of contrast injec
tion, number of contrast injections, total data acquisition time, imag
e quality, radiation dose and patient comfort. There was a reduction i
n the total procedure time, improvement in patient comfort, data acqui
sition time and a reduction in the number of injections. Conventional
DSA may underestimate stenoses in 65% of cases, There is no significan
t change in the total volume of contrast injected. There is, however,
an increase in the radiation dose during data acquisition.