A POSITIVE DISPLACEMENT MICROPUMP FOR MICRODIALYSIS

Citation
J. Cunneen et al., A POSITIVE DISPLACEMENT MICROPUMP FOR MICRODIALYSIS, Mechatronics, 8(5), 1998, pp. 561-583
Citations number
38
Categorie Soggetti
Robotics & Automatic Control","Computer Science Artificial Intelligence","Engineering, Eletrical & Electronic","Engineering, Mechanical","Robotics & Automatic Control","Computer Science Artificial Intelligence
Journal title
ISSN journal
09574158
Volume
8
Issue
5
Year of publication
1998
Pages
561 - 583
Database
ISI
SICI code
0957-4158(1998)8:5<561:APDMFM>2.0.ZU;2-4
Abstract
Miniature fluid pumps, measuring 15 x 4 x 1 mm, have been microfabrica ted with silicon, glass, and polyimide. Pumps have been tested with de ionized water and 10% glycerol solutions as the working solutions. The pumps have been operated with a pneumatic drive or a piezoelectric dr ive. Flow rates from 0.1 to 110 mu l min(-1) have been achieved. The m aximum pressure generated by the pumps was 56 cm of water. Pumps have been operated with a dialysis probe and with a microchannel load. The pump lifetime is limited by the degradation in the performance of the polyimide components in the pump. The power consumption was less than 1 mW at a drive frequency of 10 Hz. (C) 1998 Elsevier Science Ltd. All rights reserved.