ETAAS DETERMINATION OF ALUMINUM AND COPPER IN DIALYSIS CONCENTRATES AFTER MICROCOLUMN CHELATING ION-EXCHANGE PRECONCENTRATION

Citation
S. Knezevic et al., ETAAS DETERMINATION OF ALUMINUM AND COPPER IN DIALYSIS CONCENTRATES AFTER MICROCOLUMN CHELATING ION-EXCHANGE PRECONCENTRATION, Fresenius' journal of analytical chemistry, 362(1), 1998, pp. 162-166
Citations number
25
Categorie Soggetti
Chemistry Analytical
ISSN journal
09370633
Volume
362
Issue
1
Year of publication
1998
Pages
162 - 166
Database
ISI
SICI code
0937-0633(1998)362:1<162:EDOAAC>2.0.ZU;2-Q
Abstract
A procedure was developed for the preconcentration and determination o f aluminium and copper in dialysis concentrates at the ng cm(-3) level . The preconcentration was achieved on microcolumns filled with Chelex -100 resin adjusted to a pH of 4.0. Five repetitive cycles of the samp le through the column ensured a sufficient contact time for quantitati ve retention of aluminium and copper ions. The retained ions were elut ed with HNO3 (0.5 mol dm(-3)). Aluminium and copper were determined in the eluate by Zeeman ETAAS using the standard addition technique. The procedure was performed under clean room conditions (class 10,000), T he reliability of the results was evaluated by recovery tests, using d ialysis concentrates spiked with aluminium and copper. The recoveries obtained ranged from 86 to 106% for aluminium and from 92 to 97% for c opper. Using the recommended procedure, the LOD of aluminium and coppe r in dialysis concentrates (preconcentration factor 2) was found to be 0.5 ng cm(-3) and 0.2 ng cm(-3), respectively.