S. Knezevic et al., ETAAS DETERMINATION OF ALUMINUM AND COPPER IN DIALYSIS CONCENTRATES AFTER MICROCOLUMN CHELATING ION-EXCHANGE PRECONCENTRATION, Fresenius' journal of analytical chemistry, 362(1), 1998, pp. 162-166
A procedure was developed for the preconcentration and determination o
f aluminium and copper in dialysis concentrates at the ng cm(-3) level
. The preconcentration was achieved on microcolumns filled with Chelex
-100 resin adjusted to a pH of 4.0. Five repetitive cycles of the samp
le through the column ensured a sufficient contact time for quantitati
ve retention of aluminium and copper ions. The retained ions were elut
ed with HNO3 (0.5 mol dm(-3)). Aluminium and copper were determined in
the eluate by Zeeman ETAAS using the standard addition technique. The
procedure was performed under clean room conditions (class 10,000), T
he reliability of the results was evaluated by recovery tests, using d
ialysis concentrates spiked with aluminium and copper. The recoveries
obtained ranged from 86 to 106% for aluminium and from 92 to 97% for c
opper. Using the recommended procedure, the LOD of aluminium and coppe
r in dialysis concentrates (preconcentration factor 2) was found to be
0.5 ng cm(-3) and 0.2 ng cm(-3), respectively.