Purpose: To report the successful application of a method to adjust a
malpositioned bifurcated stent-graft after endovascular aortic aneurys
m repair. Method and Results: A 62-year-old male patient underwent end
ovascular repair of a 5.1-cm abdominal aortic aneurysm(AAA) with a Van
guard bifurcated stent-graft. After complete deployment of the stent-g
raft, the intraoperative completion angiogram disclosed unexpected occ
lusion of the left renal artery. Intra-aortic adjustment of the bifurc
ated graft was possible with a crossover guidewire, which was pulled c
audally. The method worked perfectly to restore blood flow to the left
renal artery. The patient is well 16 months postoperatively without a
ny evidence of endoleak or graft migration; the left renal artery rema
ins open. Conclusions:A technique is demonstrated for intra-aortic rep
ositioning of a bifurcated stent-graft to correct insufficient deploym
ent. If required, this technique should be attempted before conversion
to an open procedure.