R. Perezolmos et al., FLOW-INJECTION ANALYSIS OF FLUORIDE IN ELECTROPLATING BATHS WITH POTENTIOMETRIC DETECTION, Fresenius' journal of analytical chemistry, 362(2), 1998, pp. 230-233
A high dispersion FIA system with a fluoride selective and a double ju
nction reference electrode in cascade arrangement was developed for th
e determination of the fluoride content in chromium electroplating bat
hs. The analyses of seven synthetic baths prepared in the laboratory a
nd two samples provided by the industry gave results with a mean recov
ery value of 100.5%, and a mean variation coefficient of 1.6%. These d
ata were better than those obtained by the EDTA complexometric titrati
on, the reference technique, with a mean recovery of 94.0% and a mean
variation coefficient of 7.5%. The sampling rate for the FIA manifold
was 20 samples per hour.