A. Romolini et al., BICONICALLY-TAPERED OPTICAL-FIBER PROBES FOR THE MEASUREMENT OF ESOPHAGEAL PRESSURE, Sensors and actuators. A, Physical, 70(3), 1998, pp. 205-210
A miniaturized optical fiber probe for measuring esophageal pressure,
that makes use of biconically-tapered fibers, has been built and chara
cterized. The functioning of this probe is based on the decrease in tr
ansmitted power in the fiber, when it is bent in its biconical part un
der the action of pressure. We made and tested several probes, using d
ifferent fibers (monomode and few-modes) and different tapering values
. Our best result was achieved with a probe which utilizes a biconical
monomode fiber that has a waist diameter of 36 mu m. The resolution o
btained was 2 mm Hg in the range between 5 and 55 mm Hg. (C) 1998 Else
vier Science S.A. All rights reserved.