M. Watakabe et al., TECHNICAL ASPECTS OF MECHNOMYOGRAPHY RECORDING WITH PIEZOELECTRIC CONTACT SENSOR, Medical & biological engineering & computing, 36(5), 1998, pp. 557-561
The piezoelectric contact sensor has been widely utilised in mechanomy
ography (MMG). The authors aim to clarify the mechanical variables (i.
e. acceleration, velocity or displacement) reflected by the MMG signal
detected with a piezoelectric contact sensor (PEC), and compare the r
esults with those obtained simultaneously by an accelerometer (ACC). T
o measure the acceleration-frequency response, a mechanical sinusoidal
excitation of 5 to 300 Hz at a constant magnitude of 0.01 G was appli
ed to the two transducers. The acceleration-frequency response of the
ACC transducer was confirmed to be almost flat. The PEC without any re
striction of the transducer housing (including the combined seismic ma
ss) demonstrated a similar response to the ACC transducer. The PEC tra
nsducer output with restricted housing decreased with increasing sinus
oidal frequency and an attenuation slope of -40 dB/decade and phase an
gle of -180 degrees. The voluntary MMG signal during isometric knee ex
tension was recorded simultaneously with the two transducers. The ampl
itude spectral density distribution of the MMG from the PEC transducer
was narrow and the mean frequency was approximately one-half that obt
ained from the ACC transducer. The amplitude spectral density distribu
tion with the PEC transducer resembled that of the double integral ove
r time of the ACC transducer signal. The phase angle of the PEC transd
ucer signal was different from that of the ACC transducer signal by ap
proximately -180 degrees. These results suggest that the PEC transduce
r acts as a displacement meter of muscle vibration. In addition, diffe
rences in the MMG frequency components relating to the transducer type
must be taken into consideration when investigating the mechanical ac
tivity of muscle.