RUNAWAY PACEMAKER IN AN IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR

Citation
Cm. Carpenter et al., RUNAWAY PACEMAKER IN AN IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR, Journal of cardiovascular electrophysiology, 9(9), 1998, pp. 1008-1011
Citations number
9
Categorie Soggetti
Cardiac & Cardiovascular System
ISSN journal
10453873
Volume
9
Issue
9
Year of publication
1998
Pages
1008 - 1011
Database
ISI
SICI code
1045-3873(1998)9:9<1008:RPIAIC>2.0.ZU;2-3
Abstract
Introduction: Runaway pacemaker is a potentially catastrophic complica tion of any permanent pacing system, Methods and Results: A 70-year-ol d man was found to have erratic behavior of his implantable cardiovert er defibrillator (ICD) during a routine outpatient interrogation. His device was turned off, and he was hospitalized in preparation for a pu lse generator replacement. During his hospitalization, his ICD unexpec tedly began pacing rapidly, Despite prompt resuscitation attempts, the patient died, Postmortem examination of the device demonstrated a cry stal oscillator failure, Conclusion: A previously unrecognized compone nt malfunction is a potentially lethal complication of ICDs.