Aj. Oosterkamp et al., GRADIENT REVERSED-PHASE LIQUID-CHROMATOGRAPHY COUPLED ONLINE TO RECEPTOR-AFFINITY DETECTION BASED ON THE UROKINASE RECEPTOR, Journal of chromatography B. Biomedical sciences and applications, 715(1), 1998, pp. 331-338
Citations number
20
Categorie Soggetti
Chemistry Analytical","Biochemical Research Methods
Journal title
Journal of chromatography B. Biomedical sciences and applications
A postcolumn receptor-affinity detection (RAD) was developed for the d
etection of urokinase and cross-reactive compounds. The analytical met
hod consisted of gradient reversed-phase HPLC coupled on-line to a RAD
system based on fluorescein-labelled urokinase receptor (fluorescein-
uPAR) as reagent, Fluorescein-uPAR was added continuously to the HPLC
effluent to react with analytes eluting from the LC column. Unreacted
fluorescein-uPAR was removed by a short affinity column packed with an
immobilised urokinase support. The analyte-bound fluorescein-uPAR fra
ction passes the affinity column unretained and was detected downstrea
m by means of a fluorescence detector. An absolute detection limit of
40 fmol urokinase was obtained in the flow injection mode. In the grad
ient HPLC-RAD system a detection limit of 40 nM (20-mu l injection, ab
solute amount, 800 fmol) was obtained. The present method allowed the
identification of active breakdown products of urokinase both in stand
ard samples and biological matrices. (C) 1998 Elsevier Science B.V. Al
l rights reserved.