ACCURACY OF A PORTABLE INTERNATIONAL NORMALIZATION RATIO MONITOR IN OUTPATIENTS RECEIVING LONG-TERM ORAL ANTICOAGULANT-THERAPY - COMPARISONWITH A LABORATORY REFERENCE-STANDARD USING CLINICALLY RELEVANT CRITERIA FOR AGREEMENT
Jd. Douketis et al., ACCURACY OF A PORTABLE INTERNATIONAL NORMALIZATION RATIO MONITOR IN OUTPATIENTS RECEIVING LONG-TERM ORAL ANTICOAGULANT-THERAPY - COMPARISONWITH A LABORATORY REFERENCE-STANDARD USING CLINICALLY RELEVANT CRITERIA FOR AGREEMENT, Thrombosis research, 92(1), 1998, pp. 11-17
The accuracy of a new, portable INR monitor (CoaguChek, Boehringer-Man
nheim, Indianapolis, IN) was evaluated by comparing INR results from t
he portable monitor with results obtained by a laboratory-based method
. Dual INR measurements (portable monitor, laboratory) were performed
in 163 consecutive outpatients receiving warfarin. Agreement in dual I
NR measurements was defined based on clinically-relevant expanded and
narrow criteria and statistical criteria. Agreement in dual INR measur
ements also was evaluated as a function of increasing INR. The proport
ion of dual INR measurements that satisfied the clinically-relevant ex
panded, and narrow agree ment criteria was 90%, and 86%, respectively.
Seventy-nine percent of all dual measurements were within 0.5 INR uni
ts. The accuracy of the portable monitor was greatest for INR values l
ess than 3.0; above this INR level, the portable monitor underestimate
d laboratory INR values. The proportion of dual INR measurements withi
n 0.5 INR units for laboratory INR ranges of <2.0, 2.0-3.0, 3.1-4.0, a
nd >4.0 was 98%, 87%, 57%, and 21%, respectively. We conclude that the
portable INR monitor achieved a clinically acceptable level of accura
cy when compared to the traditional laboratory method and provides a s
uitable alternative method of monitoring the INR in patients receiving
warfarin. (C) 1998 Elsevier Science Ltd.